St Jude Medical Professional Published Articles 
 
Skip Navigation Links.

 Combined Use of Morphology Discrimination, Sudden Onset, and Stability as Discriminating Algorithms in Single Chamber Cardioverter Defibrillators 

Abstract

Morphology Discrimination (MD) is a rhythm discriminator based on QRS morphology analysis that can be combined with other discriminators like Stability, with or without Sinus Interval History (SIH) and Sudden Onset. Thirty-five patients implanted with a St. Jude Medical single chamber ICD were evaluated during exercise testing, during induced AF, and during follow-up for 14 ± 5 months. At exercise testing (60 episodes detected) MD had a specificity (SP) of 96.7% and Sudden Onset a SP of 91.7%; during induced AF (25 episodes) both MD and Stability had a SP of 96.0%. The diagnostic performance on spontaneous arrhythmias was as follows: for ventricular tachycardia (126 episodes) a sensitivity (SE) of 94.4% for MD, 92.1% for Sudden Onset, 89.7% for Stability without SIH and 79.4% for Stability + SIH; for sinus tachycardia (44 episodes) a SP of 86.4% for MD, 97.7% for Sudden Onset, 2.3% for Stability and of 95.5% for Stability + SIH. For AF (165 cases) a SP of 67.9% for MD, 69.1% for Stability and 90.3% for Stability + SIH, 44.8% for Sudden Onset. Use of MD alone provided a SE of 94.4% and a SP of 71.4% for spontaneous arrhythmias and combined use of the discriminators in a "2 of 3" diagnostic logic implied a SP of 90.9% with maintenance of 96.0% of SE. In single chamber ICDs a wide range of SE/SP ratios may be obtained by use of multiple discriminators, but use of the algorithm in a 2 of 3 diagnostic logic may achieve a SP of 90.9% and a SE of 96.0%.

Author

Giuseppe Boriani, Eraldo Occhetta, Gianfranco Pistis, Carlo Menozzi, Marcella Jorfida, Sergio Sermasi, Massimo Pagani, Gianni Gasparini, Giacomo Musso, Alberto Dall'acqua, Mauro Biffi, and Angelo Branzi

Citation

Pacing and Clinical Electrophysiology, 25(9): 357-1366(10), 1 September 2002.